strain gage animation

Japanese

"HVF" / Small & Flash-Diaphragm

hvf type Pressure Transducers

Features

  • Small size. Flush diaphragm type.
  • Most suitable for the measurement of high response impulsive pressures.
  • Covering applications for high viscosity pressure media.
  • Accuracy class: 0.5% RO.

Appearance Dimensions

product_dimension

Model and Capacity / Frequency response / etc. (unit : mm)
Model Capacity Natural frequency Weight
HVF-2MPA
2MPa(20.39kgf/cm2)
71kHz
38g
HVF-5MPA
5MPa(50.99kgf/cm2)
110kHz
38g
HVF-10MPA
10MPa(102.0kgf/cm2)
150kHz
38g
HVF-20MPA
20MPa(203.9kgf/cm2)
210kHz
38g
HVF-50MPA
50MPa(509.9kgf/cm2)
320kHz
38g

Specifications

Safe Overload   120%RC
Rated Output   1.5mV/V±20%
Nonlinearity   0.5%RO
Hysteresis   0.5%RO
Repeatability   0.3%RO
Excitation Voltage   2V (or less)
Safe Excitation Voltage   3V
Input Rsistance   120Ω
Output Rsistance   120Ω
Compensated Temp.Range   0 to 50℃
Safe Temp.Range   -10 to 60℃
Temp.Effect on Zero   0.08%RO/℃
Temp.Effect on Output   0.05%/℃
Cable   Φ4mm-4wire shielded cable, length : 0.3m
Connector   Jack / PRC03-32A10-7F
Attached cable   L-A-5  (Refer to transducer connections)